Process Viscometer KV100-Ex-TM25
The viscometer KV100-Ex works on the principle of a Capillary Viscometer
with continual volume flow and measurement of pressure decrease over the measuring
capillary.
Its basic components are the driving head and the flow measuring device connected
by a spacer tube. The driving head holds a synchronous motor with change gear. The
immersion-type measuring unit holds a gear pump, the intake fitting and the
exchangeable measuring capillary, as well as ducts for the supply of the process
pressure and for the transport of overpressure from the entrance of the measuring
capillary to the connections of pressure transducers.
The gear pump, which is driven by the driving head through a shaft via magnetic
coupling at constant speed, continually takes a constant volume flow from the process
medium and feeds it through the measuring capillary as a steady flow.
The pressure decrease between the two ends of the measuring capillary, which is
proportional to viscosity, is investigated by a differential pressure transducer.
Pressure differences are transformed into standard electrical signals 4...20 mA,
which can be used for display and as a control signal.
For direct installation on containers, tanks, or reactors at any position
the viscometer has a DIN flange connection. The inlet of the aspiration tube
must always be located within the representative fluid.
The process viscometer can be completed with an electronics unit for more
functions of instrument control, signal processing, correction of characteristic
lines of the gear pump, as well as for the temperature adjustment of viscosity.
Recommended Applications
The process viscometer KV100-Ex-TM25 is suited for continual viscosity
measurement and monitoring in a wide variety of liquid substances, such as
- lubricating oil
- polymer solutions
- fuel oil
- emulsions
- fat melts
- suspensions
- printing inks
- liquid detergents
- latex
- adhesives
- lacquers
- glues
The rheology consultants of RheoTec will be pleased to help you analysing
your specific application problem, and, based on their profound rheological
knowledge, they will find exactly the equipment configuration
you need to solve it.
Technical Data
| viscosity ranges |
0 to 3 mPas
0 to 10 mPas
0 to 30 mPas
0 to 50 mPas
0 to 100 mPas
0 to 500 mPas |
| accuracy |
± 1 % of maximum value |
| pressure range |
0 to 2.5 MPa |
| temperature range |
-25 °C to +140 °C |
| materials with direct fluid contact |
1.4571 according to DIN 17440 |
| explosion protection |
II 2G EEx de IIB T4
(according to ATEX standard) |
| degree of protection |
IP54 |
| climatic classification |
-25/55/30/90//42 (S02)1-3 |
| installation position |
any |
| power supply |
motor:
230 V AC (+6% / -10% // 50 or 60 Hz)
pressure difference transducer:
24 V DC (+10% / -20% // 20 mA) |
| data processing |
pressure difference transducer
(Δp max. 0.05 MPa) |
| output signal |
4 to 20 mA (intrinsically safe) |
Optional Accessories
1. Temperature sensor Pt100
- optional delivered with transducer
- 3 temperature ranges -30°C to +60°C, 0°C to +100°C, 0°C to +200°C
2. Control unit KS98
- provides supply voltage 24 V DC for differential pressure transducer
- shows data on display (viscosity, temperature, min./max. viscosity, compensated viscosity)
- temperature correction of viscosity data according to specified viscosity-temperature equation via large temperature range
| power supply: |
90 to 250 V AC / 48 to 62 Hz |
| protection: |
according to DIN VDE 0470 / EN 60529 |
| device panel: |
IP65 |
| housing: |
IP20 |
| connections: |
IP00 |
 |
| Control unit KS98 |
|
|
|
|