Process Viscometer KV100-25-TM25
The viscometer KV100-25 works on the principle of a Capillary Viscometer with
continual volume flow and measurement of pressure decrease over the measuring
capillary.
Its basic components are the driving head and the flow measuring device connected
by a spacer tube. The driving head holds a synchronous motor with change gear. The
immersion-type measuring unit holds a gear pump, the intake fitting and the
exchangeable measuring capillary, as well as ducts for the supply of the process
pressure and for the transport of overpressure from the entrance of the measuring
capillary to the connections of pressure transducers.
The gear pump, which is driven by the driving head through a shaft via magnetic
coupling at constant speed, continually takes a constant volume flow from the
process medium and feeds it through the measuring capillary as a steady flow.
The pressure decrease between the two ends of the measuring capillary, which is
proportional to viscosity, is investigated by a differential pressure transducer.
Pressure differences are transformed into standard electrical signals 4...20 mA,
which can be used for display and as a control signal.
For direct installation on containers, tanks, or reactors at any position the
viscometer has a DIN flange connection. The inlet of the aspiration tube must
always be located within the representative fluid.
The process viscometer can be completed with an electronics unit for more functions
of instrument control, signal processing, correction of characteristic lines of the
gear pump, as well as for the temperature adjustment of viscosity.
Recommended Applications
The process viscometer KV100-25-TM25 is suited for continual viscosity measurement
and monitoring in a wide variety of liquid substances, such as
- lubricating oil
- polymer solutions
- fuel oil
- emulsions
- fat melts
- suspensions
- printing inks
- liquid detergents
- latex
- adhesives
- lacquers
- glues
The rheology consultants of RheoTec will be pleased to help
you analysing your specific application problem, and, based on their
profound rheological knowledge, they will find exactly the equipment
configuration you need to solve it.
Technical Data
| viscosity ranges |
0 to 3 mPas
0 to 10 mPas
0 to 30 mPas
0 to 50 mPas
0 to 100 mPas
0 to 500 mPas |
| accuracy |
± 1 % of maximum value |
| pressure range |
0 to 2.5 MPa |
| temperature range |
-25 °C to +140 °C |
| materials with direct fluid contact |
1.4571 according to DIN 17440 |
| degree of protection |
IP54 |
| climatic classification |
-25/55/30/90//42 (S02)1-3 |
| installation position |
any |
| power supply |
motor:
230 V AC (+6% / -10% // 50 or 60 Hz)
pressure difference transducer:
24 V DC (+10% / -20% // 20 mA) |
| data processing |
pressure difference transducer
(Δp max. 0,05 MPa) |
| output signal |
4 to 20 mA |
Optional Accessories
1. Temperature sensor Pt100
- optional delivered with transducer
- 3 temperature ranges -30°C to +60°C, 0°C to +100°C, 0°C to +200°C
2. Control unit KS98
- provides supply voltage 24 V DC for differential pressure transducer
- shows data on display (viscosity, temperature, min./max. viscosity, compensated viscosity)
- temperature correction of viscosity data according to specified viscosity-temperature equation via large temperature range
| power supply: |
90 to 250 V AC / 48 to 62 Hz |
| protection: |
according to DIN VDE 0470 / EN 60529 |
| device panel: |
IP65 |
| housing: |
IP20 |
| connections: |
IP00 |
 |
| Control Unit KS98 |
|
|
|
|